Download PDF

Diabetes Care

Publication date: 2013-01-01
Volume: 36 Pages: 188 - 194
Publisher: American Diabetes Association

Author:

Van Herpe, Tom
Mesotten, Dieter ; Wouters, Pieter ; Herbots, J ; Voets, E ; Buyens, J ; De Moor, Bart ; Van den Berghe, Greet

Keywords:

SISTA, Science & Technology, Life Sciences & Biomedicine, Endocrinology & Metabolism, PREDICTIVE CONTROL ALGORITHM, LATE PARENTERAL-NUTRITION, INTENSIVE-CARE-UNIT, ILL PATIENTS, GLYCEMIC CONTROL, PROTOCOL, THERAPY, ICU, ASSOCIATION, MANAGEMENT, Aged, Algorithms, Blood Glucose, Critical Illness, Female, Humans, Hypoglycemic Agents, Insulin, Male, Middle Aged, 11 Medical and Health Sciences, 32 Biomedical and clinical sciences, 42 Health sciences

Abstract:

OBJECTIVE: Tight blood glucose control (TGC) in critically ill patients is difficult and labor intensive, resulting in poor efficacy of glycemic control and increased hypoglycemia rate. The LOGIC-Insulin computerized algorithm has been developed to assist nurses in titrating insulin to maintain blood glucose levels at 80-110 mg/dL (normoglycemia) and to avoid severe hypoglycemia (<40 mg/dL). The objective was to validate clinically LOGIC-Insulin relative to TGC by experienced nurses. RESEARCH DESIGN AND METHODS: The investigator-initiated LOGIC-1 study was a prospective, parallel-group, randomized, controlled clinical trial in a single tertiary referral center. A heterogeneous mix of 300 critically ill patients were randomized, by concealed computer allocation, to either nurse-directed glycemic control (Nurse-C) or algorithm-guided glycemic control (LOGIC-C). Glycemic penalty index (GPI), a measure that penalizes both hypoglycemic and hyperglycemic deviations from normoglycemia, was the efficacy outcome measure, and incidence of severe hypoglycemia (<40 mg/dL) was the safety outcome measure. RESULTS: Baseline characteristics of 151 Nurse-C patients and 149 LOGIC-C patients and study times did not differ. The GPI decreased from 12.4 (interquartile range 8.2-18.5) in Nurse-C to 9.8 (6.0-14.5) in LOGIC-C (P < 0.0001). The proportion of study time in target range was 68.6 ± 16.7% for LOGIC-C patients versus 60.1 ± 18.8% for Nurse-C patients (P = 0.00016). The proportion of severe hypoglycemic events was decreased in the LOGIC-C group (Nurse-C 0.13%, LOGIC-C 0%; P = 0.015) but not when considered as a proportion of patients (Nurse-C 3.3%, LOGIC-C 0%; P = 0.060). Sampling interval was 2.2 ± 0.4 h in the LOGIC-C group versus 2.5 ± 0.5 h in the Nurse-C group (P < 0.0001). CONCLUSIONS: Compared with expert nurses, LOGIC-Insulin improved efficacy of TGC without increasing rate of hypoglycemia.