This item still needs to be validated !
Title: Use of solid sampling zeeman atomic-absorption spectrometry (ss-zaas) for certification purposes
Authors: Hofmann, C ×
Pauwels, Jan
Vandecasteele, Carlo #
Issue Date: Aug-1994
Publisher: Springer verlag
Series Title: Fresenius journal of analytical chemistry vol:349 issue:12 pages:779-783
Abstract: Three methods for the calibration of solid sampling Zeeman atomic absorption spectrometry allowing independent calibration based on gravimetric principles and thus usable for certification purposes are described and intercompared: standard addition to the solid sample, use of synthetic solid matrix matching samples and extrapolation to zero matrix. lead to the best results, extrapolation to zero matrix is less suited at high concentrations, but is useful where an accuracy of the order of 10% is acceptable.
ISSN: 0937-0633
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Department of Kinesiology - miscellaneous