ITEM METADATA RECORD
Title: Concept, design and fabrication of smart orthopedic implants
Authors: Burny, F ×
Donkerwolcke, M
Moulart, F
Bourgois, R
Puers, Robert
Van Schuylenbergh, K
Barbosa, M
Paiva, O
Rodes, F
Begueret, JB
Lawes, P #
Issue Date: Sep-2000
Publisher: Elsevier sci ltd
Series Title: Medical engineering & physics vol:22 issue:7 pages:469-479
Abstract: Extensive clinical experience has been built up using orthopaedic implants instrumented with strain gauges connected to a Wheatstone bridge by means of percutaneous leads. This research showed the medical relevance of the monitoring of the deformation of implants as a powerful tool to evaluate nursing and rehabilitation exercises, for tracing dangerous overloads and anticipating implant failure and also to observe the healing process. The IMPACT 3500 project focuses on the instrumentation of femoral implants with on board sensors: regular Benoist-Girard implants have been modified. to contain a 'sensing cell', and thoroughly tested in vitro and in vivo. The implant deformations are measured with resistive strain gauges, and the signal is transferred to a personal computer for processing and display, via a hard wired connection, or via a telemetry system. Two fully implantable wireless designs, called Linkstrain and Sealstrain, are powered from the outside by magnetic induction. As Sealstrain contains the whole telemetric system in its cavity, the highest miniaturization was required; this seriously deteriorates the efficiency of the inductive power link. (C) 2001 IPEM. Published by Elsevier Science Ltd. All rights reserved.
URI: 
ISSN: 1350-4533
Publication status: published
KU Leuven publication type: IT
Appears in Collections:ESAT - MICAS, Microelectronics and Sensors
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy

 




All items in Lirias are protected by copyright, with all rights reserved.

© Web of science