ITEM METADATA RECORD
Title: Knee kinetics and kinematics: What are the effects of TKA malconfigurations?
Authors: Pianigiani, Silvia ×
Labey, Luc
Pascale, Walter
Innocenti, Bernardo #
Issue Date: Aug-2016
Publisher: Springer International
Series Title: Knee Surgery, Sports Traumatology, Arthroscopy vol:24 issue:8
Abstract: Purpose
Total knee arthroplasty (TKA) is a very successful surgical procedure. However, implant failures and patient dissatisfaction still persist. Sometimes surgeons are not able to understand and explain these negative performances because the patient’s medical images “look good”, but the patient “feels bad”. Apart from radiograph imaging and clinical outcome scores, conventionally used follow-up methods are mainly based on the analysis of knee kinematics. However, even if kinematics remains close to the “normal” range of motion, the patient may still complain about pain and functional limitations. To provide more insight into this paradox, a better quantitative understanding of TKA mechanics must be developed. For this purpose, improved techniques for clinical follow-up, combining kinetics and kinematics analysis, should be introduced to help surgeons to assess and understand TKA performance.

Methods
An analysis on four TKA designs was performed, and the changes in kinematics and in kinetics induced by several implant configurations (simulating implant malalignment and different knee anatomy) were compared. More specifically, analysed tibio-femoral and patello-femoral contact forces and tibio-femoral kinematics were analysed during a squat task up to 120°.

Results
The results from this study show that contact forces (with changes up to 67 %) are more heavily affected by malconfigurations than kinematics, for which maximum deviations are of the order of 5 mm or 5°, similar to the simulated surgical errors. The results present a similar trend for the different designs.

Conclusions
The results confirm the hypothesis that kinematics is not the only and also not the most relevant parameter to predict or explain knee function after TKA. In the future, techniques to analyse knee kinetics should be integrated in the clinical follow-up.
ISSN: 0942-2056
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Mechanical Engineering Technology TC, Technology Campus Geel
Biomechanics Section
× corresponding author
# (joint) last author

Files in This Item:
File Description Status SizeFormat
art%3A10.1007%2Fs00167-015-3514-y.pdf Accepted 545KbAdobe PDFView/Open Request a copy

These files are only available to some KU Leuven Association staff members

 




All items in Lirias are protected by copyright, with all rights reserved.

© Web of science