Title: Comparison of signal to noise ratios from spatial and frequency domain formulations of nonprewhitening model observers in digital mammography
Authors: Sisini, Francesco ×
Zanca, Federica #
Marshall, Nicholas W
Taibi, Angelo
Cardarelli, Paolo
Bosmans, Hilde #
Issue Date: Sep-2012
Publisher: Published for the American Association of Physicists in Medicine by the American Institute of Physics
Series Title: Medical Physics vol:39 issue:9 pages:5652-5663
Article number: 10.1118/1.4747267
Abstract: Image quality indices based upon model observers are promising alternatives to laborious human readings of contrast-detail images. This is especially appealing in digital mammography as limiting values for contrast thresholds determine, according to some international protocols, the acceptability of these systems in the radiological practice. The objective of the present study was to compare the signal to noise ratios (SNR) obtained with two nonprewhitening matched filter model observer approaches, one in the spatial domain and the other in the frequency domain, and with both of them worked out for disks as present in the CDMAM phantom.
ISSN: 0094-2405
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Medical Physics & Quality Assessment (+)