Title: Completion of a Truncated Attenuation Image from the Attenuated PET Emission Data
Authors: Nuyts, Johan ×
Bal, G
Kehren, F
Fenchel, M
Michel, C
Watson, C #
Issue Date: Feb-2013
Publisher: Institute of Electrical and Electronics Engineers
Series Title: IEEE transactions on medical imaging vol:32 issue:2 pages:237-246
Abstract: Positron emission tomographs (PET) are currently almost exclusively designed as hybrid systems. The current standard is the PET/CT combination, while prototype PET/MRI systems are being studied by several research groups. One problem in these systems is that the transaxial field of view of the second system is smaller than that of the PET camera and does not provide complete attenuation data. Because this second system provides the image for PET attenuation and scatter correction, the smaller field of view causes truncation of the attenuation map, producing bias in the attenuation corrected activity image. In this paper, we propose a maximum-a-posteriori algorithm for estimating the missing part of the attenuation map from the PET emission data. The method is evaluated on five artificially truncated 18F-FDG PET/CT studies, where it reduced the error on the reconstructed PET activities from 20% to less than 7%. The results on a PET/MRI patient study with 18F-FDG are presented as well.