ITEM METADATA RECORD
Title: Radiographic analysis of a bone plug in 275 primary cemented total hip arthroplasties
Authors: Vanderstappen, Jan ×
Simon, Jean-Pierre
Bellemans, Johan #
Issue Date: Jun-2012
Publisher: Acta Medica Belgica
Series Title: Acta Orthopaedica Belgica vol:78 issue:3 pages:350-6
Abstract: Various designs of cement restrictors for total hip arthroplasty have shown a variability in resistance to intramedullary pressure and migration. The performance of a conical bone plug was studied in 275 cemented total hip arthroplasties with a followup between 5 and 19 years. In a radiographic analysis the relation between the plug position and the cement mantle quality was investigated. The results were compared to the performance of other cement restrictors as reported in the literature. Good to excellent cement mantles were observed in 80% of the femurs. A significant correlation between the stability of the plug and the quality of the cement mantle was found. Forty-nine percent of the plugs were within 1 to 3 centimetres of the Exeter stem hollow centraliser. We also demonstrated that bone plugs performed almost always better than polyethylene or gelatine restrictors reported in other studies.
ISSN: 0001-6462
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Biomechanics, -implants and Tissue Engineering (-)
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy

 




All items in Lirias are protected by copyright, with all rights reserved.

© Web of science