Title: Are Short Columns Always The Best Option?
Authors: Desmet, Gert ×
Cabooter, Deirdre #
Issue Date: Feb-2009
Publisher: Advanstar Communications
Series Title: LC GC Europe vol:22 issue:2 pages:70-+
Abstract: Using a fixed length-variant of the kinetic plot method, it is illustrated how an analysis that is performed near the optimal flow-rate of a given commercial column can, in many cases, be performed between 50-200% faster by switching to a longer column and operating it at a higher pressure - at least, if the available instrument pressure admits so. The present article aims to show that short columns are not always the best choice to get the fastest separation.