Title: An ex vivo study on radiofrequency tissue ablation: increased lesion size by using an "expandable-wet" electrode
Authors: Miao, Y
Ni, Yicheng ×
Yu, Jie
Zhang, H
Baert, Albert
Marchal, Guy #
Issue Date: Jan-2001
Series Title: European radiology vol:11 issue:9 pages:1841-7
Abstract: The present comparative study was conducted to validate a newly developed "expandable-wet" electrode for an increased lesion size of radiofrequency ablation (RFA) on excised beef liver. The expandable-wet electrode, which allows interstitial hypertonic saline infusion through retractable curved needles, was compared with "expanded-dry" and "unexpanded-wet" electrodes for RFA lesion size and other parameters. A total of 120 lesions were created under 50 W (groups A-C) and 90 W (groups A'-C') power control mode for 10 min at each ablation site with the following groups: group A and A' of expanded-dry electrode (needles deployed but saline uninfused); group B and B' of unexpanded-wet electrode (saline infused but needle undeployed); and group C and C' of expanded-wet electrode (needles deployed and saline infused). Together with lower impedance and higher power output, the lesion size in group C (5.3+/-0.4 cm) and C' (6.0+/-1.0 cm) were significantly larger (P<0.01) than that in group A (3.3+/-0.3 cm) and A' (2.0+/-0.2 cm), and group B (3.8+/-1.0 cm) and B' (2.6+/-0.4 cm). The RFA lesion size can be significantly enlarged when the expandable electrode is complemented with interstitial hypertonic saline infusion. This design may improve the efficacy of RF tumor ablation.
ISSN: 0938-7994
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Radiology
Theragnostic Laboratory (+)
Faculty of Medicine - miscellaneous
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy


All items in Lirias are protected by copyright, with all rights reserved.

© Web of science