ITEM METADATA RECORD
Title: Improved delivery efficiency for step and shoot intensity modulated radiotherapy using a fast-tuning magnetron
Authors: Budgell, GJ ×
Martens, C
Claus, Filip #
Issue Date: Nov-2001
Publisher: IOP Pub.
Series Title: Physics in Medicine and Biology vol:46 issue:11 pages:N253-N261
Abstract: The delivery efficiency of step and shoot intensity modulated radiotherapy (IMRT) has been improved by the installation of fast-tuning magnetrons into three travelling wave linear accelerators. The IMRT delivery efficiency and the beam start-up performance have been compared before and after installation. Start-up and inter sub-field times were reduced by an average of 3.0 s. A typical start-up time from depression of the start button to beam on is now around 4 s. Delivery efficiency for a variety of clinical and quality control prescriptions was improved by an average of 30.7% (range 7.4-60.9%), depending on a complex combination of the number of sub-fields, distance moved by leaves and dose rate. For the oldest accelerator (7 years old), dosimetric accuracy was significantly improved for low dose sub-fields. The dose output was within 2% for a 1 monitor unit (MU) sub-field and 1% for a 2 MU sub-field. The two newer accelerators displayed similar or better dose characteristics even before fast-tuning, magnetron installation. Beam symmetries and flatnesses were acceptable at all energies and dose rates, and showed no obvious degradation in low dose sub-fields. It is recommended that fast-tuning magnetrons are adopted for accelerators of this design performing step and shoot IMRT.
ISSN: 0031-9155
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Non-KU Leuven Association publications
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy

 




All items in Lirias are protected by copyright, with all rights reserved.

© Web of science