Abstract: The software driving the Cerec CAD/CAM system has recently been updated. In this study, the marginal accuracy of computer-machined porcelain inlays was determined using the original and the updated software. Two types of MOD cavities in Frasaco resin teeth were prepared with either sharp or rounded box corners. The distance of the occlusal and proximal marginal interface was determined using a measuring microscope. The mean occlusal interfacial distance was 52 microns for both programs. Only the box corners showed a significantly larger space, of which the mean value was approximately 200 microns. The updated version improved the marginal accuracy at the box corners. Further reduction of the interfacial distance was accomplished by rounding the sharp box corners.