ITEM METADATA RECORD
Title: Hemolysis with the F21 hemopump
Authors: Wiebalck, A ×
Flameng, Willem #
Issue Date: Jul-1995
Series Title: The International journal of artificial organs vol:18 issue:7 pages:365-71
Abstract: The aim of this study was to determine the influence of mechanical factors on free hemoglobin using an F21 hemopump cannula. These factors (flow, rotary speed, reduction of the cross section area of the silicone tube and foreign material on the screw) were tested in a mock circulation with ox blood. Irrelevant increase in free hemoglobin were induced by foreign bodies either in the inlet or on the screw. The hemopump flow was easily affected and almost suppressed by small amounts of foreign bodies on the screw. Other reasons for elevated concentrations of plasma free hemoglobin may imply a general disorder of the pump, anticoagulation, renal function, red cells or a combination of these factors.
ISSN: 0391-3988
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Experimental Cardiac Surgery
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy

 




All items in Lirias are protected by copyright, with all rights reserved.