ITEM METADATA RECORD
Title: Experimental study of thrombogenicity and foreign body reaction induced by heparin-coated coronary stents
Authors: De Scheerder, Ivan ×
Wang, K
Wilczek, K
Meuleman, D
Van Amsterdam, R
Vogel, G
Piessens, Jan
Van de Werf, Frans #
Issue Date: Mar-1997
Series Title: Circulation vol:95 issue:6 pages:1549-53
Abstract: BACKGROUND: Results of recent randomized clinical trials have revealed a significant reduction in angiographic restenosis rate when adjunctive stenting was performed after conventional coronary balloon angioplasty. The thrombogenicity of metal stents, however, remains a concern. In the present study, we compare the thrombogenicity of heparin-coated coronary stents with that of bare metallic coronary stents. METHODS AND RESULTS: Thrombogenicity of metallic coronary stents (four heparin-coated and eight bare stents) was studied in a rat arteriovenous shunt model with the use of 125I-labeled fibrinogen and 51Cr-labeled platelets. Total clot weight after 30-minute follow-up was significantly lower in the heparin-coated stents compared with the bare stents (8.1 +/- 3.7 versus 25.8 +/- 4.6 mg; P < .001). Relative 125I and 51Cr activities in the stents were significantly higher in the bare stents than in the heparin-coated stents (125I, 1.03 +/- 0.43 versus 0.18 +/- 0.04, P = .003; 51Cr, 17.5 +/- 6.8 versus 4.4 +/- 1.0, P = .004). Subsequently, heparin-coated and bare stents were randomly implanted in the right coronary artery of 20 domestic pigs. Angiographic parameters were similar between both groups at baseline and after 6-week follow-up. Morphometry also did not show a significant difference in lumen area (bare, 1.03 +/- 0.83 mm2; heparin-coated, 1.12 +/- 0.73 mm2; P = NS) or neointimal hyperplasia (bare, 1.01 +/- 0.81 mm2; heparin-coated, 1.21 +/- 0.57 mm2; P = NS). CONCLUSIONS: Heparin coating of metallic coronary stents decreases their thrombogenicity but does not improve late vessel patency and neointimal hyperplasia at follow-up in a porcine coronary model.
URI: 
ISSN: 0009-7322
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Cardiology
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy

 




All items in Lirias are protected by copyright, with all rights reserved.

© Web of science