Title: In vivo comparison between two tip pressure transducer systems
Authors: Aubert, A E ×
Vrolix, M
De Geest, H
Van de Werf, Frans #
Issue Date: May-1995
Series Title: International journal of clinical monitoring and computing vol:12 issue:2 pages:77-83
Abstract: Experimental findings are presented of an in vivo comparison between a Sentron catheter and another tip transducer manometer: a Millar microtip catheter. Both catheters have been used simultaneously in the left ventricle of dogs. Pressure variations were elicited by drug infusion. Pressure values and derivatives obtained from both systems were compared. A cross correlation between episodes of the two pressures was computed. Results from this study showed good correlation between left ventricular systolic pressure measured with both manometers (R = 0.992, p < 0.0001), end-diastolic pressure (R = 0.809, p < 0.0001) and between first derivatives: positive derivative (R = 0.993, p < 0.0001) and negative (R = 0.634, p < 0.0001). The mean cross correlation between both pressure signals was 0.61 +/- 0.04. In the frequency domain no statistical difference was found between the location of the maxima of the peaks. It is concluded that a Sentron manometer can be a valid alternative, at a reasonable price, to a cheaper, though less accurate fluid filled catheter and a more expensive 'golden standard' microtip catheter.
ISSN: 0167-9945
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Cardiology
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy


All items in Lirias are protected by copyright, with all rights reserved.

© Web of science