ITEM METADATA RECORD
Title: A semi-automated 2D/3D marker-based registration algorithm modelling prostate shrinkage during radiotherapy for prostate cancer
Authors: Budiharto, Tom ×
Slagmolen, Pieter
Hermans, Jeroen
Maes, Frederik
Verstraete, Jan
Van den Heuvel, Frank
Depuydt, Tom
Oyen, Raymond
Haustermans, Karin #
Issue Date: Mar-2009
Publisher: Elsevier Science Publishers
Series Title: Radiotherapy and Oncology vol:90 issue:3 pages:331-336
Abstract: BACKGROUND AND PURPOSE: Currently, most available patient alignment tools based on implanted markers use manual marker matching and rigid registration transformations to measure the needed translational shifts. To quantify the particular effect of prostate gland shrinkage, implanted gold markers were tracked during a course of radiotherapy including an isotropic scaling factor to model prostate shrinkage. MATERIALS AND METHODS: Eight patients with prostate cancer had gold markers implanted transrectally and seven were treated with (neo) adjuvant androgen deprivation therapy. After patient alignment to skin tattoos, orthogonal electronic portal images (EPIs) were taken. A semi-automated 2D/3D marker-based registration was performed to calculate the necessary couch shifts. The registration consists of a rigid transformation combined with an isotropic scaling to model prostate shrinkage. RESULTS: The inclusion of an isotropic shrinkage model in the registration algorithm cancelled the corresponding increase in registration error. The mean scaling factor was 0.89+/-0.09. For all but two patients, a decrease of the isotropic scaling factor during treatment was observed. However, there was almost no difference in the translation offset between the manual matching of the EPIs to the digitally reconstructed radiographs and the semi-automated 2D/3D registration. A decrease in the intermarker distance was found correlating with prostate shrinkage rather than with random marker migration. CONCLUSIONS: Inclusion of shrinkage in the registration process reduces registration errors during a course of radiotherapy. Nevertheless, this did not lead to a clinically significant change in the proposed table translations when compared to translations obtained with manual marker matching without a scaling correction.
Description: Budiharto T., Slagmolen P., Hermans J., Maes F., Verstraete J., Van den Heuvel F., Depuydt T., Oyen R., Haustermans K., ''A semi-automated 2D/3D marker-based registration algorithm modelling prostate shrinkage during radiotherapy for prostate cancer'', Radiotherapy and oncology, vol. 90, no. 3, pp. 331-336, March 2009.
URI: 
ISSN: 0167-8140
Publication status: published
KU Leuven publication type: IT
Appears in Collections:ESAT - PSI, Processing Speech and Images
Laboratory of Experimental Radiotherapy
Radiology
× corresponding author
# (joint) last author

Files in This Item:
File Description Status SizeFormat
2573.pdf Published 572KbAdobe PDFView/Open Request a copy

These files are only available to some KU Leuven Association staff members

 




All items in Lirias are protected by copyright, with all rights reserved.