Title: Reliability of the Nidek NT-1000 non contact tonometer
Authors: Van de Velde, T ×
Zeyen, Thierry #
Issue Date: 1995
Series Title: Bulletin de la Société belge d'ophtalmologie vol:255 pages:19-22
Abstract: In this prospective study, we compared the intraocular pressure (IOP) readings of 100 patients measured with the Goldmann applanation tonometer and the Nidek NT-1000 pneumotonometer. The correlation coefficient between the Goldmann and Nidek readings was 0.86. On the average the pneumotonometer overestimated the intraocular pressure with 0.43 mm Hg. The Nidek NT-1000 non contact tonometer can be used for screening purposes provided an appropriately low IOP value is used to indicate the need for further assessment with the Goldman applanation tonometer.
ISSN: 0081-0746
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Research Group Ophthalmology