Title: Fluorescence Lifetime Standards for Time and Frequency Domain Fluorescence Spectroscopy
Authors: Boens, Noël ×
Qin, Wenwu
Basaric, Nikola
Hofkens, Johan
Ameloot, Marcel
Pouget, Jacques
Lefevre, Jean-Pierre
Valeur, Bernard
Gratton, Enrico
vandeVen, Martin
Silva, Norberto D
Engelborghs, Yves
Willaert, Katrien
Sillen, Alain
Rumbles, Garry
Phillips, David
Visser, Antonie J. W. G
van Hoek, Arie
Lakowicz, Joseph R
Malak, Henryk
Gryczynski, Ignacy
Szabo, Arthur G
Krajcarski, Don T
Tamai, Naoto
Miura, Atsushi
Pouget, Jacques #
Issue Date: Mar-2007
Series Title: Analytical Chemistry vol:79 issue:5 pages:2137-2149
Abstract: A series of fluorophores with single-exponential fluorescence decays in liquid solution at 20 degrees C were measured independently by nine laboratories using single-photon timing and multifrequency phase and modulation fluorometry instruments with lasers as excitation source. The dyes that can serve as fluorescence lifetime standards for time-domain and frequency-domain measurements are all commercially available, are photostable under the conditions of the measurements, and are soluble in solvents of spectroscopic quality (methanol, cyclohexane, water). These lifetime standards are anthracene, 9-cyanoanthracene, 9,10-diphenylanthracene, N-methylcarbazole, coumarin 153, erythrosin B, N-acetyl-L-tryptophanamide, 1,4-bis(5-phenyloxazol-2-yl)benzene, 2,5-diphenyloxazole, rhodamine B, rubrene, N-(3-sulfopropyl)acridinium, and 1,4-diphenylbenzene. At 20 degrees C, the fluorescence lifetimes vary from 89 ps to 31.2 ns, depending on fluorescent dye and solvent, which is a useful range for modern pico- and nanosecond time-domain or mega- to gigahertz frequency-domain instrumentation. The decay times are independent of the excitation and emission wavelengths. Dependent on the structure of the dye and the solvent, the excitation wavelengths used range from 284 to 575 nm, the emission from 330 to 630 nm. These lifetime standards may be used to either calibrate or test the resolution of time- and frequency-domain instrumentation or as reference compounds to eliminate the color effect in photomultiplier tubes. Statistical analyses by means of two-sample charts indicate that there is no laboratory bias in the lifetime determinations. Moreover, statistical tests show that there is an excellent correlation between the lifetimes estimated by the time-domain and frequency-domain fluorometries. Comprehensive tables compiling the results for 20 (fluorescence lifetime standard/solvent) combinations are given.
URI: 
ISSN: 0003-2700
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Molecular Imaging and Photonics
Biochemistry, Molecular and Structural Biology Section
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy

 




All items in Lirias are protected by copyright, with all rights reserved.

© Web of science