Title: Evaluation of impulse oscillation system: comparison with forced oscillation technique and body plethysmography
Authors: Hellinckx, J ×
Cauberghs, M
De Boeck, Christiane
Demedts, Maurice #
Issue Date: Oct-2001
Series Title: European Respiratory Journal vol:18 issue:3 pages:564-70
Abstract: The impulse oscillation system (IOS) has been developed recently to measure respiratory system resistance (Rrs) and reactance (Xrs) at different frequencies up to > or = 25 Hz. IOS has, however, not been validated against established techniques. This study compared IOS with the classical pseudorandom noise forced oscillation technique (FOT) and body plethysmographic airway resistance (Raw) in 49 subjects with a variety of lung disorders and a wide range of Raw (0.10-1.28 kPa x L(-1) x s). Rrs,IOS was slightly greater than Rrs,FOT, especially at lower frequencies, with a mean +/- SD difference at 5-6 Hz of 0.14 +/- 0.09 kPa x L(-1) x s. Comparisons with the wave-tube technique applied on two analogues indicated an overestimation by IOS. Xrs,IOS and Xrs,FOT were very similar, with a slightly higher resonant frequency with IOS than with FOT (mean difference +/- SD 1.35 +/- 3.40 Hz). Raw was only moderately correlated with Rrn,FOT and Rrs-IOS; although the mean differences were small (0.04 +/- 0.14 kPa x L(-1)s for Rrs6,FOT and -0.10 +/- 0.14 kPa x L(-1) x s for Rrs5,IOS), IOS and FOT markedly underestimated high resistance values. In conclusion, the impulse oscillation system yields respiratory system resistance and reactance values similar, but not identical to those provided by the forced oscillation technique.
ISSN: 0903-1936
Publication status: published
KU Leuven publication type: IT
Appears in Collections:Pneumology
Pediatric Pulmonology Section (-)
× corresponding author
# (joint) last author

Files in This Item:

There are no files associated with this item.

Request a copy


All items in Lirias are protected by copyright, with all rights reserved.

© Web of science